TRANSPORTIX PLUS

The portable system TRANSPORTIX PLUS TXL-PLUS4 is a completely integrated system, that combines a modern High Frequency X-ray Generator’s Technology with and excellent flexibility for any clinic, emergency room, Intensive Care Unit, etc. In addition, it can be also used as a external mobile unit for emergency, rural locations and difficult access places.

Category:
  • Description
  • Specifications 4 kW
  • Specifications 8 kW

Product Description

The portable system TRANSPORTIX PLUS TXL-PLUS4 is a completely integrated system, that combines a modern High Frequency X-ray Generator’s Technology with and excellent flexibility for any clinic, emergency room, Intensive Care Unit, etc. In addition, it can be also used as a external mobile unit for emergency, rural locations and difficult access places. Main features

  • Safe and easy operation
  • Constant potential high frequency operating on single phase lines
  • Radiographic operation from the X-Ray Unit Overlay Control Panel
  • Tube protection circuitry prolongs Tube life and increases system performance.
  • Equipped with closed loop control of X-Ray Tube current, kVp and filaments, which minimize potential errors and the need for readjustments
  • Standard electric outlet operation from 100 to 240 VAC for 4 kW
  • Single-Phase Line Automatic Regulation 220 / 240¦10% VAC (50 / 60Hz) for 8 kW.
  • Automatic line voltage compensation due to closed loop operation of X-ray Tube current and kVp
  • Heat Unit storage for the X-ray Tube.

CONFIGURATION:

  • High Frequency Generator 4 kW, 125 kVp, 100 mA, 250 mAs
  • Trolley
  • X-ray Tube, 40 KHU, 0.6-1.5 mm
  • Manual CollimatorCHARACTERISTICS:
    • Console control integrated on the system with digital indicator of the selection for kVp and mAs parameters, large/small focus selection, light indicator and acoustic of the X-ray exposure. It also includes a magnetic thermal limitation switch
    • Maximum Power nominal power 4 KW (@ 0,1s)
      • 125 kVp @ 32 mA
      • 100 kVp @ 40 mA
      • 50 kVp @ 80 mA
      • 40 kVp @ 100 mA
    • Auto Diagnostic System with error code, to make easy maintenance
    • Range of Kvp, from 40 to 125 (in 1 kVp steps)
    • Kvp Accuracy ± (3% +1kV)
    • Range of mA, from 5 to 100 mA, 5, 6.4, 8, 10, 12.5, 16, 20, 25, 32, 40, 50, 64, 80, 100
    • Automatic selection of the mAs, In addition, it can be shown in the display pressing simultaneously any bottom, for focus or mAs selection
    • Range of mAs, from 0.1 - 250 mAs (in 25% steps according to R’10 series)
    • mAs Accuracy ±( 5% + 0.1 mAs)
    • Range of exposure times, from 0,001 to 10 seconds, Automatic selection of the mAs. It can be shown in the display pressing simultaneously any bottom, for focus or mAs selection. (in 25% steps)
    • ms Accuracy ±(2% + 0.1 ms)
    • Fix anode tube with double focus:
      • Small focus, 0.6 mm
      • Large focus, 1.5 mm
      • Anode Angle 14º
      • Anode Heat Capacity 40,000 H.U.
      • Total (tube+colimador) Inherent Filtration 2.9 mm Al @ 75 KVp
    • Manual Collimator Centering Light Indicator. SID Measuring Tape
    • Alimentation cable 6 m.
    • X-Ray Manipulador Cable, 3 m.

    AUTOMATIC LINE POWER DETECTION SYSTEM By means of this System, the Unit detects the maximum operative Power Line adapting the Exposure Parameters to the Power available and avoiding undesired line breakdowns when operating with poor electricity lines.

CONFIGURATION:

  • High Frequency Generator 8 kW, 125 kVp, 100 mA, 250 mAs
  • Trolley
  • X-ray Tube, 40 KHU, 0.6-2.8 mm
  • Manual CollimatorCHARACTERISTICS:
    • Console control integrated on the system with digital indicator of the selection for kVp and mAs parameters, large/small focus selection, light indicator and acoustic of the X-ray exposure. It also includes a magnetic thermal limitation switch
    • Maximum Power nominal power 8 KW (@ 0,1s)
      • 125 kVp @ 64 mA
      • 100 kVp @ 80 mA
      • 80 kVp @ 100 mA
    • Auto Diagnostic System with error code, to make easy maintenance
    • Range of Kvp, from 40 to 125 (in 1 kVp steps)
    • Kvp Accuracy ± (3% +1kV)
    • Range of mA, from 5 to 100 mA, 5, 6.4, 8, 10, 12.5, 16, 20, 25, 32, 40, 50, 64, 80, 100
    • Automatic selection of the mAs, In addition, it can be shown in the display pressing simultaneously any bottom, for focus or mAs selection
    • Range of mAs, from 0.1 - 250 mAs (in 25% steps according to R’10 series)
    • mAs Accuracy ±( 5% + 0.1 mAs)
    • Range of exposure times, from 0,001 to 10 seconds, Automatic selection of the mAs. It can be shown in the display pressing simultaneously any bottom, for focus or mAs selection. (in 25% steps)
    • ms Accuracy ±(2% + 0.1 ms)
    • Fix anode tube with double focus:
      • Small focus, 0.6 mm
      • Large focus, 2.8 mm
      • Anode Angle 14º
      • Anode Heat Capacity 76,000 H.U.
      • Total (tube+colimador) Inherent Filtration 2.9 mm Al @ 75 KVp
    • Manual Collimator Centering Light Indicator. SID Measuring Tape
    • Alimentation cable 6 m.
    • X-Ray Manipulador Cable, 3 m.

    AUTOMATIC LINE POWER DETECTION SYSTEM By means of this System, the Unit detects the maximum operative Power Line adapting the Exposure Parameters to the Power available and avoiding undesired line breakdowns when operating with poor electricity lines.